Description
GM550
The GM550 features proprietary auto-coding technology, a backlit display for reading in the dark, and other advanced functions. This mid-range model provides users with accurate results and a hassle-free experience for optimal diabetes management.
Features:
- Fast Results
- Strip Design
- Backlit Screen
- Auto-Coding
- PC Connection
Specifications
Test strip | GS550 | |
---|---|---|
Coding | Auto-coding | |
Enzyme / Measurement technology | GOD / Electrochemical sensor | |
Blood sample source | Capillary whole blood | |
Minimum sample volume | 0.75 µL | |
HCT range | 30 – 57% | |
Measuring range | 0.6 – 33.3 mmol/L (10 – 600 mg/dL) | |
Meter dimensions (L x W x H) | 90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch) | |
Display area (L x W) | 46.6 x 33 mm (1.83 x 1.32 inch) | |
Weight (with battery) | 53.0 ± 5 g | |
Power supply | 2 × CR2032 batteries | |
Memory capacity | 500 test results with date and time | |
Operating temperature | 10 – 40°C (50 – 104°F) | |
Operating relative humidity | 10 – 90% | |
Meter storage conditions | -10 – 60°C (14 – 140°F) | |
Test strip storage conditions | At 4 – 30°C (39 – 86°F), and below 90% relative humidity | |
Accuracy | For glucose concentrations < 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements. For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 15% of reference measurements. | |